Data Foundry

Medical device recalls 2024

ADVIA Centaur CA 19-9 — Class II medical device recall Z-3103-2024

Ongoing recall by Siemens Healthcare Diagnostics, Inc., reported 2024-09-18, distributed nationwide. Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous

Recall numberZ-3103-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., East Walpole, MA, United States
Recall initiated2024-07-26
Classified2024-09-11
Reported by FDA2024-09-18
DistributedNationwide (US)
Reason classespotency
UPC / GTIN / UDI-DI00630414574257
Lot numbers55970535, 55971535

Product

ADVIA Centaur CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491379 SMN: 11206239

Reason for recall

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.