Data Foundry

Medical device recalls 2024

Medline convenience kits containing semi-rigid suction liners labeled… — Class II medical device recall Z-3117-2024

Ongoing recall by MEDLINE INDUSTRIES, LP - Northfield, reported 2024-09-18, distributed nationwide. Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500

Recall numberZ-3117-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMEDLINE INDUSTRIES, LP - Northfield, Northfield, IL, United States
Recall initiated2024-06-27
Classified2024-09-11
Reported by FDA2024-09-18
DistributedNationwide (US)
Countries outside the USCA, PA
Quantity134 units
UPC / GTIN / UDI-DI10195327077990, 40195327077991
Lot numbers23LMA774, 23LME672, 24AMA991, 24CMB554, 24CMB648

Product

Medline convenience kits containing semi-rigid suction liners labeled as follows: KIT NEURO SHUNT ST LOUIS CHILD, Pack Number DYNJ907829B

Reason for recall

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3117-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.