Data Foundry

Medical device recalls 2024

Blood Pressure Module BPM3 TM-2917B — Class II medical device recall Z-3124-2024

Ongoing recall by Fresenius Medical Care Holdings, Inc., reported 2024-09-18, distributed nationwide. Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005

Recall numberZ-3124-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmFresenius Medical Care Holdings, Inc., Waltham, MA, United States
Recall initiated2024-08-08
Classified2024-09-11
Reported by FDA2024-09-18
DistributedNationwide (US)
Countries outside the USCA
Quantity3,718 systems
Reason classessoftware
UPC / GTIN / UDI-DI00840861100897, 00840861100903, 00840861100910, 00840861100927, 00840861100934, 00840861102082, 00840861102099, 00840861102105, 00840861102112
Serial numbers00840861100903, 00840861100910, 00840861100927, 00840861100934, 00840861102082, 00840861102099, 00840861102105, 00840861102112, 190766, 190858, 190895, 190908, 191124, 191126, 191128, 191130, 4T0S275391, 4T0S275399, 4T0S275410, 4T0S275417, 4T0S275422, 4T0S275440, 4T0S275444, 4T0S275451, 4T0S275455 and 475 more
Model numbers190573, 190713, 190766, 190858, 190895, 190908, 191124, 191126, 191128, 191130, 191378, 191389

Product

Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 Series Hemodialysis Machines with Part Numbers 190573, 190713, 190766, 190858, 190895, 190908, 191124, 191126, 191128, and 191130

Reason for recall

Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3124-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.