Data Foundry

Medical device recalls 2026

StatStrip Glucose and B- Ketone meter — Class I medical device recall Z-3130-2026

Ongoing recall by Nova Biomedical Corporation, reported 2026-09-23, distributed nationwide. A software synchronization/timing issue in the barcode scanning and patient data processing component that cau

Recall numberZ-3130-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmNova Biomedical Corporation, Waltham, MA, United States
Recall initiated2026-07-29
Classified2026-09-11
Reported by FDA2026-09-23
DistributedNationwide (US)
Countries outside the USAT, BE, BO, BR, CA, CH, CL, DE, FI, FR, GB, HR, IE, IL, IT, KR, KY, NL, NZ, TH, TR
Quantity1,858 units
Reason classesdevice malfunction, software
UPC / GTIN / UDI-DI00385480636834
Model numbers63683

Product

StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.

Reason for recall

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.