Data Foundry

Medical device recalls 2026

Medline Convenience Kit: OPEN HEART A PACK — Class II medical device recall Z-3137-2026

Ongoing recall by Medline Industries, LP, reported 2026-09-23, distributed nationwide. The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile ba

Recall numberZ-3137-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2026-07-23
Classified2026-09-10
Reported by FDA2026-09-23
DistributedNationwide (US)
Quantity135 units
Reason classesforeign material, sterility, packaging
UPC / GTIN / UDI-DI10198459445941, 40198459445942
Lot numbers25HLA462, 25HLA916, 25JLA847, 25LLA298, 26CLA372

Product

Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N

Reason for recall

The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.