Data Foundry

Medical device recalls 2026

INCA — Class I medical device recall Z-3139-2026

Ongoing recall by CooperSurgical, Inc., reported 2026-09-30, distributed nationwide. Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or

Recall numberZ-3139-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCooperSurgical, Inc., Trumbull, CT, United States
Recall initiated2026-08-11
Classified2026-09-18
Reported by FDA2026-09-30
DistributedNationwide (US)
Countries outside the USBR, PE
UPC / GTIN / UDI-DI20888937009433, 20888937009457, 20888937009464, 20888937009471, 60888937009448
Lot numbers550019722, 550031611, 550031614, 550031615, 550031616, 550031617, 550031618, 550031619, 550031620, 550031621, 550033779, 550033780, 550033781, 550034760, 550034763
Model numbers44-2707, 44-2709, 44-2710, 44-2712, 44-2715

Product

INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.

Reason for recall

Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.