Automated Impella Controller — Class I medical device recall Z-3148-2026
Ongoing recall by Abiomed, Inc., reported 2026-09-23, distributed nationwide. The potential for purge cassette recognition issues due to the failure of the purge flag component of the purg
| Recall number | Z-3148-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abiomed, Inc., Danvers, MA, United States |
| Recall initiated | 2026-08-12 |
| Classified | 2026-09-16 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BE, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, ES, FI, FR, GB, GR, HK, HR, HU, IE, IL, IN, IT, JP, KW, LU, MX, MY, NL, NO, NZ, PA, PK, PL, RS, SA, SE, SG, SI, TH, TW |
| UPC / GTIN / UDI-DI | 00813502010022, 00813502010428, 00813502011289, 00813502011296, 00813502011418 |
| Serial numbers | IC1004, IC1006, IC1007, IC1008, IC1011, IC1014, IC1015, IC1017, IC1019, IC1024, IC1025, IC1026, IC1027, IC1028, IC1029, IC1030, IC1031, IC1032, IC1033, IC1035, IC1036, IC1037, IC1038, IC1039, IC1040 and 475 more |
Product
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN. 4. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, Packaged, US; Product Code: 0042-0000-US.
Reason for recall
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3148-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.