Impella Optical Controller — Class I medical device recall Z-3150-2026
Ongoing recall by Abiomed, Inc., reported 2026-09-23, distributed nationwide. The potential for purge cassette recognition issues due to the failure of the purge flag component of the purg
| Recall number | Z-3150-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abiomed, Inc., Danvers, MA, United States |
| Recall initiated | 2026-08-12 |
| Classified | 2026-09-16 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AT, AU, BE, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, ES, FI, FR, GB, GR, HK, HR, HU, IE, IL, IN, IT, JP, KW, LU, MX, MY, NL, NO, NZ, PA, PK, PL, RS, SA, SE, SG, SI, TH, TW |
| UPC / GTIN / UDI-DI | 00813502011401, 00813502011814, 00813502012217, 00813502012484, 00813502012958, 00813502012972, 00813502013160 |
| Serial numbers | IC10000, IC10001, IC10002, IC10003, IC10004, IC10005, IC10006, IC10007, IC10008, IC10009, IC10010, IC10011, IC10016, IC10018, IC10019, IC10020, IC10021, IC10022, IC10025, IC10026, IC10027, IC10028, IC10029, IC10030, IC10031 and 475 more |
Product
Impella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU. 2. Optical, AIC, Impella Connect, Packaged, CA; Product Code: 0042-0040-CA. 3. Optical, AIC, Impella Connect, Packaged, EU; Product Code: 0042-0040-EU. 4. Optical, AIC, Impella Connect, Packaged, JP; Product Code: 0042-0040-JP. 5. Optical, AIC, Impella Connect, Packaged, UK; Product Code: 0042-0040-UK. 6. Optical, AIC, Impella Connect, Packaged, US; Product Code: 0042-0040-US. 7. Optical, AIC, Impella Connect, Packaged, KR; Product Code: 0042-0040-KR;
Reason for recall
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.