Philips Medical Systems Technologies Ltd — Class II medical device recall Z-3155-2026
Ongoing recall by Philips North America Llc, reported 2026-10-07. The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to
| Recall number | Z-3155-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | FDA mandated |
| Recalling firm | Philips North America Llc, Cambridge, MA, United States |
| Recall initiated | 2026-08-18 |
| Classified | 2026-09-29 |
| Reported by FDA | 2026-10-07 |
| Quantity | 62 |
| UPC / GTIN / UDI-DI | 00884838101111 |
Product
Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333
Reason for recall
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3155-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.