Data Foundry

Medical device recalls 2024

The device is a mobile x-ray system designed to work with Fujifilm's… — Class II medical device recall Z-3158-2024

Ongoing recall by FUJIFILM Healthcare Americas Corporation, reported 2024-10-02, distributed nationwide. When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx

Recall numberZ-3158-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedFDA mandated
Recalling firmFUJIFILM Healthcare Americas Corporation, Lexington, MA, United States
Recall initiated2023-11-23
Classified2024-09-25
Reported by FDA2024-10-02
DistributedNationwide (US)
Quantity130

Product

The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

Reason for recall

When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3158-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.