IV Sets with drip chambers — Class II medical device recall Z-3160-2026
Ongoing recall by ICU Medical, Inc., reported 2026-09-23, distributed nationwide. Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic mol
| Recall number | Z-3160-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | ICU Medical, Inc., San Clemente, CA, United States |
| Recall initiated | 2026-05-12 |
| Classified | 2026-09-11 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Quantity | 100,712,657 |
| Reason classes | specification failure |
| UPC / GTIN / UDI-DI | 10840619024021, 10840619024069, 10840619024120, 10840619064409, 10840619099388, 10887709004112, 10887709020518, 10887709020907, 10887709022017, 10887709022215, 10887709022352, 10887709022444, 10887709022871, 10887709022987, 10887709023090, 10887709023601, 10887709024110, 10887709024257, 10887709024295, 10887709024578, 10887709024967, 10887709026886, 10887709028248, 10887709028965, 10887709028972 and 64 more |
Product
IV Sets with drip chambers
Reason for recall
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3160-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.