Data Foundry

Medical device recalls 2024

System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236… — Class II medical device recall Z-3165-2024

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2024-09-25, distributed nationwide. Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts

Recall numberZ-3165-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands
Recall initiated2024-08-22
Classified2024-09-18
Reported by FDA2024-09-25
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, KY, LB, LK, LT, LV, MA, MT, MX, MY, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA
Quantity7,091 units
Reason classesdevice malfunction
UPC / GTIN / UDI-DI00884838085251, 00884838085268, 00884838085275, 00884838085282, 00884838085350, 00884838085367, 00884838099203, 00884838099210, 00884838099227, 00884838099234, 00884838099241, 00884838099258, 00884838099265, 00884838099272, 00884838110564, 00884838110571, 00884838116764, 00884838116771, 00884838116788, 00884838116801

Product

System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azurion 3 M15 722221 Azurion 3 M12 722079 Azurion 7 M20 722078 Azurion 7 M12 722068 Azurion 7 B20 722067 Azurion 7 B12 722064 Azurion 3 M15 722063 Azurion 3 M12

Reason for recall

Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3165-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.