Atellica CH Microalbumin_2 — Class II medical device recall Z-3169-2024
Ongoing recall by Siemens Healthcare Diagnostics, Inc., reported 2024-09-25, distributed nationwide. Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-D
| Recall number | Z-3169-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Tarrytown, NY, United States |
| Recall initiated | 2024-08-15 |
| Classified | 2024-09-18 |
| Reported by FDA | 2024-09-25 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BH, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GR, HR, HU, ID, IE, IL, IN, IR, IT, JO, JP, KE, KR, KW, LT, LV, MA, MX, MY, NL, NO, OM, PE, PK, PL, PT, PY, QA, RO, RS, SE, SG, SI, SK, TH, TN, TR, TW, UY, VN |
| Quantity | 18,483 units |
| UPC / GTIN / UDI-DI | 00630414596310 |
| Lot numbers | 232033, 232128, 232137, 232146, 232147, 242149, 242150, 242365 |
Product
Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers
Reason for recall
Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3169-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.