VITROS Chemistry Products Cl- Slides — Class III medical device recall Z-3184-2018

Terminated recall by Ortho-Clinical Diagnostics, reported 2018-09-26, distributed nationwide. The slides failed to meet current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quant

Recall numberZ-3184-2018
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2018-07-11
Classified2018-09-17
Reported by FDA2018-09-26
Terminated2020-08-11
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PL, PT, RU, SE, SG
Quantity168,919

Product

VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage: VITROS Cl Slides (Product Code 684 4471) - For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

Reason for recall

The slides failed to meet current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for urine samples. The measuring range (15-300 mmol/L) for urine samples when tested with the Cl- slide is not affected. The current Cl- LoQ is 5 mmol/L and the LoQ determined by this investigation is 15 mmol/L.