Data Foundry

Medical device recalls 2026

Aisys CS2 with Et Control — Class II medical device recall Z-3191-2026

Ongoing recall by Datex-Ohmeda, Inc., reported 2026-09-23, distributed nationwide. Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to

Recall numberZ-3191-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDatex-Ohmeda, Inc., Madison, WI, United States
Recall initiated2026-07-30
Classified2026-09-16
Reported by FDA2026-09-23
DistributedNationwide (US)
Quantity7 units
Serial numbersANAM00443, ANAP00609, ANAP01088, ANAQ00988, ANAR00543, ANAS00586, ANAS01357
Model numbers1011-9000-000

Product

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Reason for recall

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3191-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.