Et Control — Class II medical device recall Z-3192-2026
Ongoing recall by Datex-Ohmeda, Inc., reported 2026-09-23, distributed nationwide. Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to
| Recall number | Z-3192-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Datex-Ohmeda, Inc., Madison, WI, United States |
| Recall initiated | 2026-07-30 |
| Classified | 2026-09-16 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Quantity | 410 units |
| Serial numbers | 107841136, 107851114, 107861401, 107906045-1, 107906890, 107918076, 107919759, 107941857, 107945491, 107978351, 107978475, 108053916-1, 108053916-2, 108073330-1, 108105559, 108186429, 108189840, 108207332, 108224454-1, 108226503-1, 20033613, 20034085, 20034430-1, 20034430-2, 20034430-3 and 385 more |
| Model numbers | 1011-8000-000-S, 5700665 |
Product
Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665
Reason for recall
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3192-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.