Data Foundry

Medical device recalls 2026

Captus 4000e Thyroid Uptake System — Class II medical device recall Z-3193-2026

Ongoing recall by Mirion Technologies (Capintec), Inc., reported 2026-09-23, distributed nationwide. Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest

Recall numberZ-3193-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMirion Technologies (Capintec), Inc., Florham Park, NJ, United States
Recall initiated2026-07-20
Classified2026-09-16
Reported by FDA2026-09-23
DistributedNationwide (US)
Countries outside the USAT, BD, BR, CA, CL, CR, DZ, EC, ES, FR, GB, HK, IL, IN, IT, KR, MY, PH, PL, SA, SE, TH, TR, TW, VN
Quantity783
UPC / GTIN / UDI-DI00859942006096, 00859942006102, 00859942006423
Serial numbers940548
Model numbers5430-30151, 5430-30152, 5430-30154

Product

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

Reason for recall

Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.