Captus 4000e Thyroid Uptake System — Class II medical device recall Z-3193-2026
Ongoing recall by Mirion Technologies (Capintec), Inc., reported 2026-09-23, distributed nationwide. Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest
| Recall number | Z-3193-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Mirion Technologies (Capintec), Inc., Florham Park, NJ, United States |
| Recall initiated | 2026-07-20 |
| Classified | 2026-09-16 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | AT, BD, BR, CA, CL, CR, DZ, EC, ES, FR, GB, HK, IL, IN, IT, KR, MY, PH, PL, SA, SE, TH, TR, TW, VN |
| Quantity | 783 |
| UPC / GTIN / UDI-DI | 00859942006096, 00859942006102, 00859942006423 |
| Serial numbers | 940548 |
| Model numbers | 5430-30151, 5430-30152, 5430-30154 |
Product
Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;
Reason for recall
Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.