2008T HEMODIALYSIS SYS — Class II medical device recall Z-3194-2018
Terminated recall by Fresenius Medical Care Renal Therapies Group, LLC, reported 2018-09-26, distributed nationwide. In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interru
| Recall number | Z-3194-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Fresenius Medical Care Renal Therapies Group, LLC, Waltham, MA, United States |
| Recall initiated | 2018-07-13 |
| Classified | 2018-09-18 |
| Reported by FDA | 2018-09-26 |
| Terminated | 2019-09-17 |
| Distributed | Nationwide (US) |
| Quantity | 80 |
| Reason classes | chemical contamination |
| Model numbers | 190713, 191138 |
Product
2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.
Reason for recall
In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.