Data Foundry

Medical device recalls 2018

2008T HEMODIALYSIS SYS — Class II medical device recall Z-3194-2018

Terminated recall by Fresenius Medical Care Renal Therapies Group, LLC, reported 2018-09-26, distributed nationwide. In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interru

Recall numberZ-3194-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmFresenius Medical Care Renal Therapies Group, LLC, Waltham, MA, United States
Recall initiated2018-07-13
Classified2018-09-18
Reported by FDA2018-09-26
Terminated2019-09-17
DistributedNationwide (US)
Quantity80
Reason classeschemical contamination
Model numbers190713, 191138

Product

2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.

Reason for recall

In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3194-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.