Data Foundry

Medical device recalls 2026

AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No — Class II medical device recall Z-3195-2026

Ongoing recall by Howmedica Osteonics Corp., reported 2026-09-23, distributed nationwide. Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in

Recall numberZ-3195-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHowmedica Osteonics Corp., Mahwah, NJ, United States
Recall initiated2026-07-30
Classified2026-09-16
Reported by FDA2026-09-23
DistributedNationwide (US)
Countries outside the USCA, ZA
Quantity12,156 us
Reason classesdevice malfunction
Lot numbersP22E159
Model numbersC-M112-2-500

Product

AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Reason for recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.