Data Foundry

Medical device recalls 2026

ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512… — Class II medical device recall Z-3197-2026

Ongoing recall by Howmedica Osteonics Corp., reported 2026-09-23, distributed nationwide. Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in

Recall numberZ-3197-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHowmedica Osteonics Corp., Mahwah, NJ, United States
Recall initiated2026-07-30
Classified2026-09-16
Reported by FDA2026-09-23
DistributedNationwide (US)
Countries outside the USCA, ZA
Quantity12,156 us
Reason classesdevice malfunction
UPC / GTIN / UDI-DI04546540364319, 04546540400611, 04546540400635, 04546540400758, 07613154020515, 07613154020539, 07613154020652, 07613154020737, 07613327014082, 07613327017014, 07613327017069, 07613327017120, 07613327017168, 07613327017182, 07613327017212, 07613327017267, 07613327017281, 07613327017359, 07613327025491, 07613327033779, 07613327033847, 07613327034240, 07613327034356, 07613327034875, 07613327050288 and 3 more
Lot numbers14HDSA, 14NACD, 14SCVA, 14SCVD, 14SCVF, 17376651, 17879451, 19093651, 193307A, 193307E, 1R8TNH, 22764851, 23125951, 23325151, 23505751, 23640051, 25621551, 26862851, 27835951, 28208551, 3044-0026, 3052-1115, 30582401, 30582601, 324H9H, 414A2L, 4W1HD7, 5D4NMJ, 5J8PM2, 72-16-0512, 72-16-0518, 72-16-0912, 72-16-1112, 72-4-0324, 72-4-9114, 73-0110, 73-0910, 73-2610, 73-2710, 78-5-0910 and 43 more

Product

ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PATELLA 73-2710 SCORPIO INSET PATELLA 73-2610 SCORPIO M-DOME PATELLA 73-0110 73-0710 73-0910 SCORPIO RECESSED PATELLA 26MM 3044-0026 SCORPIO TS TIB INS 72-4-9114 SCORPIO TS TIBIAL INSERT 72-4-0324 SCORPIOFLEX TOTAL KNEE CR 72-16-0512 72-16-0518 72-16-0912 72-16-1112 SERIES II TIBIAL BEAR INSERT NO 11 15MM 3052-1115 UHR BIPOLAR 26X52mm UH1-52-26 UHR BIPOLAR 26X58mm UH1-58-26 UHR BIPOLAR 28X48mm UH1-48-28 UHR BIPOLAR 28X52mm UH1-52-28 UHR BIPOLAR 28X55mm UH1-55-28 UHR BIPOLAR 32X49MM UH1-49-32 UHR BIPOLAR 32X51MM UH1-51-32 UHR BIPOLAR 32X52MM UH1-52-32 UHR BIPOLAR 32X54MM UH1-54-32 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Reason for recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3197-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.