ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM… — Class II medical device recall Z-3201-2026
Ongoing recall by Howmedica Osteonics Corp., reported 2026-09-23, distributed nationwide. Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in
| Recall number | Z-3201-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Howmedica Osteonics Corp., Mahwah, NJ, United States |
| Recall initiated | 2026-07-30 |
| Classified | 2026-09-16 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, ZA |
| Quantity | 12,156 us |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 07613327024289, 07613327024319, 07613327024326, 07613327024357, 07613327024364, 07613327024449, 07613327025446, 07613327025453, 07613327025460, 07613327025477, 07613327025484, 07613327025538, 07613327025545, 07613327025552, 07613327025583, 07613327025590, 07613327031430, 07613327031447, 07613327031454, 07613327031591, 07613327038538, 07613327138870, 07613327138979, 07613327313581 |
| Lot numbers | 123J5R, 13SFLA, 15NKT, 191723, 191867, 194RN0, 19ECK, 1K5VVW, 2099-2252, 2099-2256, 2099-2861, 2367AK, 274HMK, 284X4X, 29385D, 2K2DW1, 3A2P68, 3H5RRK, 3M77NN, 3X0NJD, 4100R9, 4A40TV, 4W4MPL, 5J8P89, 5L82T0, 5V78KA, 620-00-28D, 6299-N-332, 6302-H-632, 6636-0-419, 69-2244, 69-2246, 69-2248, 69-2850, 69-2852, 690-00-22D, 690-00-28F, 690-00-32I, 690-10-22E, 690-10-22F and 62 more |
Product
ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INSERT 48MM 69-2248 ALL POLY CONSTRAINED INSERT 50MM 69-2850 ALL POLY CONSTRAINED INSERT 52MM 69-2852 CONSTRAINED ACETABULAR INSERTS 22MM 50MM 2099-2250 CONSTRAINED ACETABULAR INSERTS 22MM 52MM 2099-2252 CONSTRAINED ACETABULAR INSERTS 22MM 56MM 2099-2256 CONSTRAINED ACETABULAR INSERTS 28MM 61MM 2099-2861 EIUS UNI TIB SMALL RM/LL 9MM 6636-0-419 PCA CONST LINER 32MM NEUTRAL C - SHELL 6299-N-332 SYS 12 CONST LINER 32ID 10 DEG P6 +4MM 6302-H-632 TRIDENT 0 DEG CONSTRAINED INSERT 22D 690-00-22D TRIDENT 0 DEG CONSTRAINED INSERT 22E 690-00-22E TRIDENT 0 DEG CONSTRAINED INSERT 28F 690-00-28F TRIDENT 0 DEG CONSTRAINED INSERT 28G 690-00-28G TRIDENT 0 DEG CONSTRAINED INSERT 32I 690-00-32I TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D TRIDENT CONSTR ACET INSERT 690-10-28K TRIDENT CONSTRAINED INSERT 690-10-22E 690-10-22F 690-10-28G 690-10-28H 690-10-28I Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason for recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3201-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.