Data Foundry

Medical device recalls 2026

BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS… — Class II medical device recall Z-3202-2026

Ongoing recall by Howmedica Osteonics Corp., reported 2026-09-23, distributed nationwide. Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in

Recall numberZ-3202-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHowmedica Osteonics Corp., Mahwah, NJ, United States
Recall initiated2026-07-30
Classified2026-09-16
Reported by FDA2026-09-23
DistributedNationwide (US)
Countries outside the USCA, ZA
Quantity12,156 us
Reason classesdevice malfunction
UPC / GTIN / UDI-DI07613327590074
Lot numbers194593, C-M063-5-800, C-M113-5-600, MH621A, RX06358A

Product

BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Reason for recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3202-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.