SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030… — Class II medical device recall Z-3203-2026
Ongoing recall by Howmedica Osteonics Corp., reported 2026-09-23, distributed nationwide. Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in
| Recall number | Z-3203-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Howmedica Osteonics Corp., Mahwah, NJ, United States |
| Recall initiated | 2026-07-30 |
| Classified | 2026-09-16 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, ZA |
| Quantity | 12,156 us |
| Reason classes | device malfunction |
| Lot numbers | 12WAHA, 12WAHB, 13VAW, 13VAWA, 3045-0030 |
Product
SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason for recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.