Data Foundry

Medical device recalls 2026

SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110… — Class II medical device recall Z-3206-2026

Ongoing recall by Howmedica Osteonics Corp., reported 2026-09-23, distributed nationwide. Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in

Recall numberZ-3206-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHowmedica Osteonics Corp., Mahwah, NJ, United States
Recall initiated2026-07-30
Classified2026-09-16
Reported by FDA2026-09-23
DistributedNationwide (US)
Countries outside the USCA, ZA
Quantity12,156 us
Reason classesdevice malfunction
UPC / GTIN / UDI-DI07613327033847, 07613327033854, 07613327033861, 07613327033885, 07613327033892, 07613327033908, 07613327033915
Lot numbers14AGFD, 14NEYA, 1L45, 1XED, 3467, 3D8D, 3W25, 4JM5, 73-3110, 7EL5, 9Y3T, AME3, D1M9, D2YP, EX8T, HMM5, KRT3, KRT3X1, MK74, NTV7, PMRHX1, T325, T6YJ, T9ME, V8X8, X6YE

Product

SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Reason for recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3206-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.