Data Foundry

Medical device recalls 2026

CADD Yellow Medication Cassette Reservoir with NRFit connector US… — Class II medical device recall Z-3212-2026

Ongoing recall by ICU Medical, Inc., reported 2026-09-23, distributed nationwide. Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result i

Recall numberZ-3212-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmICU Medical, Inc., Lake Forest, IL, United States
Recall initiated2026-08-20
Classified2026-09-17
Reported by FDA2026-09-23
DistributedNationwide (US)
Countries outside the USAT, AU, BE, CR, DE, EE, ES, FI, FR, GB, IE, IT, JP, KW, NL, NO, SE
Quantity185,688 units
Reason classespackaging
UPC / GTIN / UDI-DI30610586044001, 30610586044018
Lot numbers21-7609-24, 6022079, 6026878, 6037697, 6037698, 6037784, 6053978, 6054015, 6061525, 6062724, 6070575, 6070616, 6072964, 6072970, 6077773, 6082332, 6092846, 6092886, 6101798, 6101906, 6108514, 6108515, 6108516, 6108556, 6126029, 6140861, 6140862, 6140865, 6140893, 6143454, 6147696, 6147697, 6155095, 6155132, 6162769, 6166032, 6171586, 6171587

Product

CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24

Reason for recall

Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3212-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-29. Not affiliated with or endorsed by FDA. Not medical or legal advice.