CADD Yellow Medication Cassette Reservoir with NRFit connector US… — Class II medical device recall Z-3212-2026
Ongoing recall by ICU Medical, Inc., reported 2026-09-23, distributed nationwide. Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result i
| Recall number | Z-3212-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | ICU Medical, Inc., Lake Forest, IL, United States |
| Recall initiated | 2026-08-20 |
| Classified | 2026-09-17 |
| Reported by FDA | 2026-09-23 |
| Distributed | Nationwide (US) |
| Countries outside the US | AT, AU, BE, CR, DE, EE, ES, FI, FR, GB, IE, IT, JP, KW, NL, NO, SE |
| Quantity | 185,688 units |
| Reason classes | packaging |
| UPC / GTIN / UDI-DI | 30610586044001, 30610586044018 |
| Lot numbers | 21-7609-24, 6022079, 6026878, 6037697, 6037698, 6037784, 6053978, 6054015, 6061525, 6062724, 6070575, 6070616, 6072964, 6072970, 6077773, 6082332, 6092846, 6092886, 6101798, 6101906, 6108514, 6108515, 6108516, 6108556, 6126029, 6140861, 6140862, 6140865, 6140893, 6143454, 6147696, 6147697, 6155095, 6155132, 6162769, 6166032, 6171586, 6171587 |
Product
CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24
Reason for recall
Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.