Data Foundry

Medical device recalls 2026

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with… — Class II medical device recall Z-3215-2026

Ongoing recall by Heidelberg Engineering GmbH, reported 2026-09-30, distributed nationwide. Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stabi

Recall numberZ-3215-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHeidelberg Engineering GmbH, Heidelberg, Germany
Recall initiated2026-08-06
Classified2026-09-18
Reported by FDA2026-09-30
DistributedNationwide (US)
Countries outside the USAT, AU, CA, CH, CR, CZ, DE, ES, FR, GB, IL, IN, IT, MX, NL, PH, PL, SA, TH, VN, ZA
Quantity87
Reason classesspecification failure
UPC / GTIN / UDI-DI04059237000235
Lot numbersSPEC-07265, SPEC-11121, SPEC-11281, SPEC-12096, SPEC-13305, SPEC-14403, SPEC-15282, SPEC-15587, SPEC-20071, SPEC-20348, SPEC-20351, SPEC-20591, SPEC-20605, SPEC-23473, SPEC-23941, SPEC-23950, SPEC-23975, SPEC-24507, SPEC-24558, SPEC-24582, SPEC-24611
Model numbersS1610, S2000, S2610, S3610

Product

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye

Reason for recall

Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3215-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.