Midmark IQvitals Zone — Class II medical device recall Z-3218-2026
Ongoing recall by Midmark Corporation, reported 2026-09-30, distributed nationwide. A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
| Recall number | Z-3218-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Midmark Corporation, Versailles, OH, United States |
| Recall initiated | 2026-08-21 |
| Classified | 2026-09-18 |
| Reported by FDA | 2026-09-30 |
| Distributed | Nationwide (US) |
| Quantity | 153 units |
| UPC / GTIN / UDI-DI | 00841709105807, 00841709105883 |
| Serial numbers | L1187570, V2838889, V2839311, V2839313, V2839314, V2840480, V2841435, V2841440, V2841442, V2841486, V2841701, V2841713, V2841724, V2841750, V2842081, V2842097, V2842105, V2842116, V2842133, V2842901, V2845499, V2845500, V2845501, V2846602, V2866067 and 281 more |
| Model numbers | 1-200-0330, 1-200-03300 |
Product
Midmark IQvitals Zone, automatically and noninvasively measures systolic and diastolic blood pressure, pulse rate, temperature (temporal), and oxygen saturation (SpO2) for adult and pediatric patients, Model Number: 1-200-03300 with Kit Number: 4-000-0624
Reason for recall
A performance issue may result in inaccurate blood pressure readings when the Linear Deflation Method is used.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3218-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.