Data Foundry

Medical device recalls 2026

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories — Class I medical device recall Z-3227-2026

Ongoing recall by Medline Industries, LP, reported 2026-10-07, distributed nationwide. Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendan

Recall numberZ-3227-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2026-08-14
Classified2026-09-30
Reported by FDA2026-10-07
DistributedNationwide (US)
Countries outside the USBM, CA, PA
Quantity771,127 total
Reason classesdevice malfunction
UPC / GTIN / UDI-DI10197344119059, 10197344119066, 10884389161692, 10888277363021

Product

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED

Reason for recall

Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.