Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories — Class I medical device recall Z-3227-2026
Ongoing recall by Medline Industries, LP, reported 2026-10-07, distributed nationwide. Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendan
| Recall number | Z-3227-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medline Industries, LP, Northfield, IL, United States |
| Recall initiated | 2026-08-14 |
| Classified | 2026-09-30 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Countries outside the US | BM, CA, PA |
| Quantity | 771,127 total |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 10197344119059, 10197344119066, 10884389161692, 10888277363021 |
Product
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED
Reason for recall
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.