Data Foundry

Medical device recalls 2026

VITEK 2 AST-N461 TEST KIT — Class II medical device recall Z-3231-2026

Ongoing recall by Biomerieux Inc, reported 2026-09-30. The issue is related to card pouch damage associated with a portion of the lot which was produced on a specifi

Recall numberZ-3231-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBiomerieux Inc, Hazelwood, MO, United States
Recall initiated2026-07-24
Classified2026-09-18
Reported by FDA2026-09-30
Countries outside the USAT, DE, LU
Quantity97,680 units
UPC / GTIN / UDI-DI03573026637712
Lot numbers1333393504

Product

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Reason for recall

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3231-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.