VITEK 2 AST-N461 TEST KIT — Class II medical device recall Z-3231-2026
Ongoing recall by Biomerieux Inc, reported 2026-09-30. The issue is related to card pouch damage associated with a portion of the lot which was produced on a specifi
| Recall number | Z-3231-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Biomerieux Inc, Hazelwood, MO, United States |
| Recall initiated | 2026-07-24 |
| Classified | 2026-09-18 |
| Reported by FDA | 2026-09-30 |
| Countries outside the US | AT, DE, LU |
| Quantity | 97,680 units |
| UPC / GTIN / UDI-DI | 03573026637712 |
| Lot numbers | 1333393504 |
Product
VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
Reason for recall
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3231-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.