Data Foundry

Medical device recalls 2017

Platinium CRT-D 1741 — Class II medical device recall Z-3233-2017

Terminated recall by Sorin Group Italia SRL - CRF, reported 2017-10-04, distributed nationwide. There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and C

Recall numberZ-3233-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSorin Group Italia SRL - CRF, Saluggia, VC, N/A, Italy
Recall initiated2017-07-24
Classified2017-09-27
Reported by FDA2017-10-04
Terminated2024-01-29
DistributedNationwide (US)
Quantity68 units
UPC / GTIN / UDI-DI08031527015484
Model numbersTDF037U

Product

Platinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).

Reason for recall

There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3233-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.