Data Foundry

Medical device recalls 2026

Evolution upgrade 1 — Class II medical device recall Z-3235-2026

Ongoing recall by Philips North America Llc, reported 2026-09-30, distributed nationwide. A software issue will terminate an examination during the image reconstruction process if the scan parameters

Recall numberZ-3235-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America Llc, Cambridge, MA, United States
Recall initiated2026-08-21
Classified2026-09-18
Reported by FDA2026-09-30
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, BZ, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, KY, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, VE, VN, ZA
Quantity91
Reason classessoftware
UPC / GTIN / UDI-DI00884838099722, 00884838099739, 00884838108714
Serial numbers41011, 41024, 41025, 41031, 41043, 41051, 41080, 41085, 41130, 41139, 41144, 41167, 41179, 41189, 41196, 41229, 41284, 41286, 41327, 41344, 41349, 41375, 41392, 41413, 41424 and 66 more
Model numbers782116, 782117, 782143, 782148, 884838099722, 884838099739, 884838108714

Product

Evolution upgrade 1.5T and Evolution upgrade 3.0T; Model Numbers (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782117 (3.0T), 4. 782143 (3.0T);

Reason for recall

A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3235-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.