Name — Class II medical device recall Z-3240-2026
Ongoing recall by Philips North America Llc, reported 2026-09-30, distributed nationwide. A software issue will terminate an examination during the image reconstruction process if the scan parameters
| Recall number | Z-3240-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips North America Llc, Cambridge, MA, United States |
| Recall initiated | 2026-08-21 |
| Classified | 2026-09-18 |
| Reported by FDA | 2026-09-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, BZ, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, KW, KY, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, VE, VN, ZA |
| Quantity | 82 |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00884838088108 |
| Serial numbers | 10403, 11009, 11029, 11039, 11040, 11043, 11044, 11090, 12001, 12046, 12049, 12057, 12075, 12078, 12079, 12081, 12086, 12099, 12105, 12106, 13006, 13012, 13017, 13048, 13049 and 57 more |
| Model numbers | 781104, 781171, 781172, 781173, 781195, 781295, 884838088108 |
Product
Name (Model Number): 1. Intera 1.5T (781195), 2. Intera 1.5T (781295), 3. Intera 1.5T Achieva IT Nova (781175), 4. Intera 1.5T Achieva Nova (781172), 5. Intera 1.5T Achieva Nova-Dual (781173), 6. Intera 1.5T Omni/Stellar (781104), 7. Intera Achieva 1.5T Pulsar (781171);
Reason for recall
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3240-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.