Data Foundry

Medical device recalls 2024

Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software… — Class II medical device recall Z-3248-2024

Ongoing recall by Hologic Inc, reported 2024-10-02, distributed nationwide. Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combinat

Recall numberZ-3248-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHologic Inc, Newark, DE, United States
Recall initiated2024-08-19
Classified2024-09-25
Reported by FDA2024-10-02
DistributedNationwide (US)
Countries outside the USBE, CA, IL, KR, NL
Quantity425 units
Reason classessoftware
UPC / GTIN / UDI-DI15420045505636, 15420045511897
Model numbersASY-16824FMI

Product

Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as follows Model/Catalog Number: 1. ASY-16824 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.1 SYSTEMS 2. ASY-16825 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.1 SYSTEMS 3. ASY-17219 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.10.0 SYSTEMS 4. ASY-17220 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.1.0 SYSTEMS 5. ASY-17221 FMI UPGRADE KIT, DIMENSIONS 1.12.0.3 SYSTEM SOFTWARE UPGRADE FOR 1.11.0 SYSTEMS 6. ASY-17222 FMI UPGRADE KIT, 3DIMENSIONS 2.3.0.3 SYSTEM SOFTWARE UPGRADE FOR 2.2.0 SYSTEMS 7. 3DM-LIC-SW-UP-V2.3 DIMENSIONS SOFTWARE 2.3 (3DIMENSIONS) 8. SDM-LIC-SW-UP-V1.12 DIMENSION SOFTWARE 1.12 (SELENIA DIMENSIONS)

Reason for recall

Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility being a false negative diagnosis

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3248-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.