Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234 — Class II medical device recall Z-3249-2026
Ongoing recall by Philips Ultrasound, LLC, reported 2026-09-30, distributed nationwide. Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measuremen
| Recall number | Z-3249-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Ultrasound, LLC, Bothell, WA, United States |
| Recall initiated | 2026-08-07 |
| Classified | 2026-09-18 |
| Reported by FDA | 2026-09-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AT, CH, DE, DK, ES, FI, FR, GB, GR, IT, JP, KR, LT, NL, NO, PH, PL, RO, SE, SI, TW, VN |
| Quantity | 219 units |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00884838102514, 00884838107540, 00884838114159, 00884838124912, 00884838128125 |
| Model numbers | 795234 |
Product
Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234
Reason for recall
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3249-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.