ForeSight Elite Tissue Oximetry Cable REF HEMFSM10 UDI-DI code:… — Class II medical device recall Z-3250-2026
Ongoing recall by Becton, Dickinson and Company, reported 2026-09-30, distributed nationwide. Due to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
| Recall number | Z-3250-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Becton, Dickinson and Company, Irvine, CA, United States |
| Recall initiated | 2026-08-11 |
| Classified | 2026-09-21 |
| Reported by FDA | 2026-09-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | KY |
| Quantity | 15,455 cables |
| Reason classes | labeling |
| UPC / GTIN / UDI-DI | 00690103210927 |
| Serial numbers | 1128702, 1907002, 1907003, 1907004, 1923033, 1924039, 1924040, 1925003, 1925004, 1925006, 1925008, 1928009, 1928010, 1928035, 1928057, 1938082, 1938085, 1938088, 1941066, 1941067, 1942006, 1944199, 1950005, 1950006, 1950008 and 475 more |
| Model numbers | 10026986001, 10026986002, 10026986003, 10026986R03, 10034992001, 10034992003, 10034992005, 10034992010, 10034992011, 10034992012, 10034992013, 10034992017, 10034992018, 10034992019, 10034992020, 10034992022, 10034992023, 10034992024, 10034992026, 10034992028, 10034992R01, 10034992R03, 10034992R10, 10034992R11, 10034992R12 and 10 more |
Product
ForeSight Elite Tissue Oximetry Cable REF HEMFSM10 UDI-DI code: 00690103210927 The non-invasive ForeSight oximeter cable is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no flow ischemic states. The ForeSight Oximeter Cable is also intended to monitor relative changes of total hemoglobin of blood under the sensors. The ForeSight Oximeter Cable is intended to allow for the display of StO2 and relative change in total hemoglobin on the HemoSphere advanced monitor.
Reason for recall
Due to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3250-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.