Automated Impella Controller — Class II medical device recall Z-3251-2026
Ongoing recall by Abiomed, Inc., reported 2026-09-30, distributed nationwide. Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical benc
| Recall number | Z-3251-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Abiomed, Inc., Danvers, MA, United States |
| Recall initiated | 2019-02-19 |
| Classified | 2026-09-21 |
| Reported by FDA | 2026-09-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AT, BE, CA, CH, DE, ES, FR, GB, HK, IT, LU, PL |
| Quantity | 3,322 units |
| Reason classes | specification failure, software |
| UPC / GTIN / UDI-DI | 00813502010022, 00813502010961, 00813502010985, 00813502011272, 00813502011289, 00813502011296, 00813502011401 |
| Serial numbers | IC1004, IC1006, IC1007, IC1008, IC1014, IC1015, IC1017, IC1025, IC1026, IC1027, IC1028, IC1029, IC1030, IC1031, IC1032, IC1039, IC1040, IC1041, IC1042, IC1043, IC1044, IC1046, IC1047, IC1049, IC1050 and 475 more |
Product
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.
Reason for recall
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3251-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-06. Not affiliated with or endorsed by FDA. Not medical or legal advice.