Data Foundry

Medical device recalls 2026

Automated Impella Controller — Class II medical device recall Z-3251-2026

Ongoing recall by Abiomed, Inc., reported 2026-09-30, distributed nationwide. Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical benc

Recall numberZ-3251-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAbiomed, Inc., Danvers, MA, United States
Recall initiated2019-02-19
Classified2026-09-21
Reported by FDA2026-09-30
DistributedNationwide (US)
Countries outside the USAT, BE, CA, CH, DE, ES, FR, GB, HK, IT, LU, PL
Quantity3,322 units
Reason classesspecification failure, software
UPC / GTIN / UDI-DI00813502010022, 00813502010961, 00813502010985, 00813502011272, 00813502011289, 00813502011296, 00813502011401
Serial numbersIC1004, IC1006, IC1007, IC1008, IC1014, IC1015, IC1017, IC1025, IC1026, IC1027, IC1028, IC1029, IC1030, IC1031, IC1032, IC1039, IC1040, IC1041, IC1042, IC1043, IC1044, IC1046, IC1047, IC1049, IC1050 and 475 more

Product

Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.

Reason for recall

Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.