Data Foundry

Medical device recalls 2026

HeartSine samaritan PAD — Class III medical device recall Z-3270-2026

Ongoing recall by HeartSine Technologies Ltd, reported 2026-10-07, distributed nationwide. Products were distributed without the required Maintenance Supplement, which describes an extra check every 3

Recall numberZ-3270-2026
Product typeMedical device
ClassificationClass III
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmHeartSine Technologies Ltd, Belfast, United Kingdom
Recall initiated2026-07-27
Classified2026-09-25
Reported by FDA2026-10-07
DistributedNationwide (US)
Quantity31,834
Serial numbers23E90139374B24, 23E90144949A24, 23G90000302H23, 23G90000303H23, 23G90000304H23, 23G90000306H23, 23G90000308H23, 23G90000312H23, 23G90000313H23, 23G90000314H23, 23G90000316H23, 23G90000369H23, 23G90000370H23, 23G90000371H23, 23G90000372H23, 23G90000373H23, 23G90000374H23, 23G90000375H23, 23G90004172G23, 23G90008119L23, 23G90009030H23, 23G90009034H23, 23G90009038H23, 23G90009047L23, 23G90009051H23 and 475 more
Model numbers350-STR-AA-AV, 350-STR-US-10, 350-STR-US-GW, 360-STR-US-10, 450-STR-US-10

Product

HeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (American Airlines) - 2 Min CPR UDI-DI code: 0506016712207 350P US Device, US English with Gateway - 2 Min CPR UDI-DI code: 0506016712804 HeartSine SAM 360P 360P US Device, US English - 2 Min CPR UDI-DI code: 0506016712068 HeartSine SAM 450P SAM 450P US Device, US English (2 min CPR) UDI-DI code: 0506016712069

Reason for recall

Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3270-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.