HeartSine samaritan PAD — Class III medical device recall Z-3270-2026
Ongoing recall by HeartSine Technologies Ltd, reported 2026-10-07, distributed nationwide. Products were distributed without the required Maintenance Supplement, which describes an extra check every 3
| Recall number | Z-3270-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class III |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | HeartSine Technologies Ltd, Belfast, United Kingdom |
| Recall initiated | 2026-07-27 |
| Classified | 2026-09-25 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Quantity | 31,834 |
| Serial numbers | 23E90139374B24, 23E90144949A24, 23G90000302H23, 23G90000303H23, 23G90000304H23, 23G90000306H23, 23G90000308H23, 23G90000312H23, 23G90000313H23, 23G90000314H23, 23G90000316H23, 23G90000369H23, 23G90000370H23, 23G90000371H23, 23G90000372H23, 23G90000373H23, 23G90000374H23, 23G90000375H23, 23G90004172G23, 23G90008119L23, 23G90009030H23, 23G90009034H23, 23G90009038H23, 23G90009047L23, 23G90009051H23 and 475 more |
| Model numbers | 350-STR-AA-AV, 350-STR-US-10, 350-STR-US-GW, 360-STR-US-10, 450-STR-US-10 |
Product
HeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US English (American Airlines) - 2 Min CPR UDI-DI code: 0506016712207 350P US Device, US English with Gateway - 2 Min CPR UDI-DI code: 0506016712804 HeartSine SAM 360P 360P US Device, US English - 2 Min CPR UDI-DI code: 0506016712068 HeartSine SAM 450P SAM 450P US Device, US English (2 min CPR) UDI-DI code: 0506016712069
Reason for recall
Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.