Positron Emission Tomography and Computed Tomography System — Class II medical device recall Z-3286-2026
Ongoing recall by Shanghai United Imaging Healthcare Co., Ltd., reported 2026-10-07, distributed nationwide. Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local
| Recall number | Z-3286-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Shanghai United Imaging Healthcare Co., Ltd., Shanghai, China |
| Recall initiated | 2026-08-11 |
| Classified | 2026-09-25 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Quantity | 3 |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 06971576832231 |
Product
Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo with software
Reason for recall
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3286-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.