Data Foundry

Medical device recalls 2026

Ingenia Ambition S and Ingenia Ambition X — Class II medical device recall Z-3310-2026

Ongoing recall by Philips North America, reported 2026-10-07, distributed nationwide. When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the e

Recall numberZ-3310-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips North America, Cambridge, MA, United States
Recall initiated2026-08-28
Classified2026-09-25
Reported by FDA2026-10-07
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BD, BE, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EC, EG, ES, FI, FR, GB, GH, GR, HK, HN, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, LB, LU, LV, MA, MT, MX, MY, NI, NL, NO, NP, NZ, PE, PH, PK, PL, PT, QA, RO, RS, SA, SE, SG, SK, TH, TN, TW, UY, VN, ZA
Quantity422
Reason classessoftware
UPC / GTIN / UDI-DI00884838090040, 00884838090057, 00884838098343, 00884838098350, 00884838108622, 00884838108639, 00884838115705, 00884838115712, 00884838128477, 00884838128484
Serial numbers22401, 22410, 29016, 29017, 29018, 29020, 29026, 29029, 29031, 29041, 29054, 29063, 29069, 29071, 29076, 29078, 29079, 29086, 29104, 29115, 29118, 29144, 29145, 29150, 29160 and 397 more
Model numbers00884838090040, 00884838090057, 00884838098343, 00884838098350, 00884838108622, 00884838108639, 00884838115705, 00884838115712, 00884838128477, 00884838128484, 781356, 781359, 782108, 782109, 782138, 782139, 782159, 782160, 782194, 782195

Product

Ingenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 782159 (S), 5. 782194 (S), 6. 781356 (X), 7. 782109 (X), 8. 782138 (X), 9. 782160 (X), 10. 782195 (X); Software Versions: R11, R12 and R13;

Reason for recall

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3310-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.