Intrapulmonary Percussive Ventilation — Class II medical device recall Z-3319-2026
Ongoing recall by Percussionaire Corporation, reported 2026-10-07, distributed nationwide. Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile a
| Recall number | Z-3319-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Percussionaire Corporation, Sandpoint, ID, United States |
| Recall initiated | 2026-08-18 |
| Classified | 2026-09-28 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AU, BE, CA, CH, CL, DE, ES, FR, GB, IL, IT, KW, LV, NL, NO, OM, PH, PT, QA, RU, SA, SE, TR, ZA |
| Quantity | 61,858 |
| Reason classes | labeling |
| UPC / GTIN / UDI-DI | 00849436000679, 00849436000686, 00849436000709 |
Product
Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual
Reason for recall
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3319-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.