Brand Name: Tandem t:slim mobile application Product Name: t:slim… — Class II medical device recall Z-3330-2026
Ongoing recall by Tandem Diabetes Care, Inc., reported 2026-10-07, distributed nationwide. A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventi
| Recall number | Z-3330-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Tandem Diabetes Care, Inc., San Diego, CA, United States |
| Recall initiated | 2026-07-16 |
| Classified | 2026-09-29 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Quantity | 177 |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00389152470641, 00389152470740 |
Product
Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
Reason for recall
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3330-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.