Systems/units containing affected software/SQL BECKMAN COULTER… — Class II medical device recall Z-3332-2026

Ongoing recall by Beckman Coulter Mishima K.K., reported 2026-10-07, distributed nationwide. Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System

Recall numberZ-3332-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Mishima K.K., Sunto-Gun, Japan
Recall initiated2026-08-03
Classified2026-09-30
Reported by FDA2026-10-07
DistributedNationwide (US)
Countries outside the USAE, AU, BG, BR, CA, CN, CR, CY, CZ, DE, DZ, EG, FR, GT, HR, HU, IE, IN, IT, JP, KR, LV, MA, MY, NG, NP, PK, PL, RO, RS, SA, TR, TW, VN, ZA
Quantity492 units
Reason classessoftware
UPC / GTIN / UDI-DI14987666540671, 14987666542965, 14987666542972
Model numbersB12183, B96692, B96693, N3151400

Product

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.

Reason for recall

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3332-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.