Systems/units containing affected software/SQL BECKMAN COULTER… — Class II medical device recall Z-3332-2026
Ongoing recall by Beckman Coulter Mishima K.K., reported 2026-10-07, distributed nationwide. Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System
| Recall number | Z-3332-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Beckman Coulter Mishima K.K., Sunto-Gun, Japan |
| Recall initiated | 2026-08-03 |
| Classified | 2026-09-30 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AU, BG, BR, CA, CN, CR, CY, CZ, DE, DZ, EG, FR, GT, HR, HU, IE, IN, IT, JP, KR, LV, MA, MY, NG, NP, PK, PL, RO, RS, SA, TR, TW, VN, ZA |
| Quantity | 492 units |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 14987666540671, 14987666542965, 14987666542972 |
| Model numbers | B12183, B96692, B96693, N3151400 |
Product
Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.
Reason for recall
Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.