Data Foundry

Medical device recalls 2026

CHARLOTTE CLAW Screw — Class II medical device recall Z-3334-2026

Ongoing recall by Wright Medical Technology, Inc., reported 2026-10-07, distributed nationwide. Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufact

Recall numberZ-3334-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmWright Medical Technology, Inc., arlington, TN, United States
Recall initiated2026-08-18
Classified2026-09-30
Reported by FDA2026-10-07
DistributedNationwide (US)
Quantity34
UPC / GTIN / UDI-DI00840420109705
Lot numbers1826666
Model numbers40122716

Product

CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System

Reason for recall

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3334-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.