CHARLOTTE CLAW Screw — Class II medical device recall Z-3334-2026
Ongoing recall by Wright Medical Technology, Inc., reported 2026-10-07, distributed nationwide. Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufact
| Recall number | Z-3334-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Wright Medical Technology, Inc., arlington, TN, United States |
| Recall initiated | 2026-08-18 |
| Classified | 2026-09-30 |
| Reported by FDA | 2026-10-07 |
| Distributed | Nationwide (US) |
| Quantity | 34 |
| UPC / GTIN / UDI-DI | 00840420109705 |
| Lot numbers | 1826666 |
| Model numbers | 40122716 |
Product
CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System
Reason for recall
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3334-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-10-13. Not affiliated with or endorsed by FDA. Not medical or legal advice.